Remote Control Commands
R&S
®
SGT100A
471
User Manual 1176.8674.02 ─ 07
Parameters:
<SBandwidth>
WIDE | NARRow
NARRow
The synchronization bandwidth is.
WIDE
Synchronization bandwidth is.
Example:
ROSC:SOUR EXT
Selects the external source.
ROSC:EXT:FREQ 10 MHz
Informs the instrument that the external reference has a fre-
quency of 10 MHz.
ROSC:EXT:SBAN WID
Selects wideband setting for synchronization bandwidth.
Manual operation:
See
[:SOURce]:ROSCillator[:INTernal]:ADJust[:STATe]
<State>
Determines whether the calibrated (
OFF
) or a user-defined (
ON
) adjustment value is
used for fine adjustment of the frequency.
If user-defined values are used, the instrument is no longer in the calibrated state.
However, the calibration value is not changed and the instrument resumes the calibra-
ted state after sending the command
:SOURce:ROSCillator:INTernal:ADJust:
STATe OFF
.
Parameters:
<State>
0 | 1 | OFF | ON
*RST:
-
Example:
ROSC:SOUR INT
Selects the internal source.
ROSC:ADJ ON
Activates use of a user-defined adjustment value.
ROSC:ADJ:VAL 1400
Sets the adjustment value to 1400.
Manual operation:
See
[:SOURce]:ROSCillator[:INTernal]:ADJust:VALue
<Value>
Allows an application to shift the reference oscillator frequency by a small amount.
The setting range depends on the reference oscillator type and its factory calibration
value. Allowed are the following ranges:
●
For TCXO oscillator: Max - Min = 1023
●
For OCXO oscillator: Max - Min = 65535 (option R&S
SGT-B1 required.)