Remote Control Commands
R&S
®
SGT100A
406
User Manual 1176.8674.02 ─ 07
[:SOURce<hw>]:BBIN:POWer:RMS?
This command queries the rms level of the external digital baseband signal.
Return values:
<Rms>
float
Range:
-100 to 10
Increment: 0.01
*RST:
0
Example:
BBIN:POW:RMS
queries the estimated rms level.
Usage:
Query only
Manual operation:
See
[:SOURce<hw>]:BBIN:SKEW
<Skew>
This command determines the delay between Q and I channel. Positive values repre-
sent a delay for Q versus I.
Parameters:
<Skew>
float
Range:
-1E-9 to 1E-9
Increment: 1E-12
*RST:
0
Default unit: s
Example:
BBIN:SKEW -23PS
enters a delay between Q and I channel of 23 picoseconds.
[:SOURce<hw>]:BBIN:SRATe[:ACTual]
<Actual>
Sets the sample rate of the external digital baseband signal.
Parameters:
<Actual>
float
Range:
400 to 200E6
Increment: 0.001
*RST:
100E6
Example:
:SOURce:BBIN:SRATe:ACTual 50E6
Manual operation:
See
[:SOURce<hw>]:BBIN:SRATe:SOURce
<Source>
This command queries the source for estimating the sample rate of the digital input sig-
nal or defining it by the user.
SOURce Subsystem