Signal Generator Settings
R&S
®
SGT100A
89
User Manual 1176.8674.02 ─ 07
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:BLANk:APPend
Level / Clock /Marker
► To access this dialog, select "SGMA-GUI > Instrument Name > Baseband > ARB >
Multi Segment.. > Segment Table".
The Level/Marker/Clock section provides the parameters necessary to adjust the
level, marker and clock settings for the selected multi segment waveform file.
Level
Defines the level mode for the multi segment waveform.
"Unchanged"
Concerning the level settings, the segments are output exactly as
defined in the files. The "Level" display applies only to the segment
with the highest rms value. In some circumstances, the remaining
segments are output at a lower level than that displayed.
"Equal RMS"
Segments are output so that all segments have the same rms value.
The "Level" display applies to all segments.
Remote command:
[:SOURce<hw>]:BB:ARBitrary:WSEGment:CONFigure:LEVel[:MODE]
Clock
Selects the clock rate mode for the multi segment waveform.
"Unchanged"
A segment is output with the clock rate defined in the file.
If segments have different clock rates, extended trigger mode "Next
Segment" allows internal segment switchovers only ("Internal" trigger
source).
Extended trigger mode "Next Segment Seamless" can only be
selected if all segments have the same clock rate.
"Highest"
All segments are output at the highest available clock rate.
This mode provides very short switchover times between segments.
The time for calculating the multi segment waveform is increased
since the individual segments have to be resampled.