Instrument setup
R&S
®
RTO6
94
User Manual 1801.6687.02 ─ 05
●
"Duplicate": the external display shows the same content as the instrument display.
●
"Extend": the instrument display and the external display show different content.
●
"Projector only": the instrument's user interface is only shown on the external dis-
play, the instrument display is off.
●
"Tablet PC settings": opens the Windows setup for touch display to adjust the touch
functionality when an additional monitor is in use.
●
"Additional settings": opens the Windows configuration for display settings.
4.4.5
Performance settings
Access: "Menu" > "Settings" > "Display" > "Performance"
Show
Displays the "Performance" results box.
The "Performance" result box shows information on the current acquisition perfor-
mance values of the R&S
RTO6.
The instrument groups acquired waveforms together in a frame, and displays the frame
content. The maximum number of frames displayed per second is about 30. The cur-
rent number of frames per second is indicated as reciprocal "Time per frame". If the
time scale decreases, and thus the number of acquisitions per second also decreases,
the number of acquisitions per frame can drop to 1.
4.5
Frontpanel settings
In the "Frontpanel" dialog box, you can assign functions to keys and knobs, adjust the
Multiuse knob and adjust the brightness of the keys.
4.5.1
Hardkeys: function assignment
Access: "Settings" > "Frontpanel" > "Hardkeys"
Frontpanel settings