Protocol analysis
R&S
®
RTO6
792
User Manual 1801.6687.02 ─ 05
Data
Sets the specified data to be searched for. The data setup consists of the condition and
one or two data patterns.
"Condition"
Defines the operator to set a specific data, e.g. "Equal" or "Not
Equal") or a range.
"Data Min/Data"
Defines the bit pattern of the data pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Data Max"
The second data pattern is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
SEARch:TRIGger:HBTO:DATA:DCONdition
Index
Sets the specified value or range within this series of data that is considered for the
search. The index setup consists of the condition and one or two index values.
"Condition"
Defines the operator to set a specific index, e.g. "Equal" or a range.
"Index Min/Index"
Defines the bit pattern of the index pattern.
In binary format, use the following characters: 1; 0; or X (do not care).
To enter the pattern in other formats use the bit pattern editor, see
Chapter 13.1.7, "Bit pattern editor"
"Index Max"
The second index pattern is required to specify a range with condi-
tions "In range".
Remote command:
SEARch:TRIGger:HBTO:DATA:ICONdition
Errors
Sets the type of error events to be searched for.
Preamble Error ← Errors
Searches for a frame with invalid preamble.
Remote command:
SEARch:TRIGger:HBTO:ERRor:PREamble
CRC Error ← Errors
Searches for a frame that has a mismatch of the Cyclic Redundancy Check (CRC)
value between the transmitting and receiving device.
Remote command:
Ethernet 100BASE-T1 (option R&S
RTO6-K560)