Protocol analysis
R&S
®
RTO6
691
User Manual 1801.6687.02 ─ 05
2. Select the tab of the bus you want to set up, for example "SB1".
3. In the "Setup" tab, select the protocol: "MIL-STD-1553".
4. Tap the "Polarity" button, and select the waveform of the data line.
5. Set the logical thresholds.
6. If necessary, tap the "Check min gap" button to select it and set the minimum gap
time.
7. If necessary, tap the "Check max response" button to select it and set the maxi-
mum response time.
13.12.3
MIL-STD-1553 trigger
13.12.3.1
Trigger settings MIL-STD-1553
Access: [Bus] > "Shortcuts" tab > "Setup trigger"
Make sure that:
●
In the [Bus] > "Setup" tab:
–
The correct "Protocol" is selected.
–
The data sources of the serial bus are channel signals.
●
In the "Trigger" > "Setup" tab:
–
The "Trigger on" = "Single event"
–
The trigger "Source" = "Serial bus".
–
The correct "Serial bus" is selected.
Trigger Type
Selects the trigger type for MIL-STD-1553 analysis.
"Sync"
Triggers on a sync impulse.
MIL-1553 (option R&S
RTO6-K530)