Protocol analysis
R&S
®
RTO6
678
User Manual 1801.6687.02 ─ 05
Remote command:
TRIGger<m>:CXPI:DATA:DCONdition
Index
The data index setup consists of the condition and one or two index patterns.
"Condition"
Defines the operator to set a specific data index ("Equal" or "Not
equal") or data index range.
"Index min /
Index"
Defines the bit pattern of the data index.
"Index max"
The second data index is required to specify a range with conditions
"In range" and "Out of range".
Remote command:
TRIGger<m>:CXPI:DATA:ICONdition
NM
Sets the value of the network management (NM) field, 2 bits indicating if the frame is in
wakeup or sleep mode.
Remote command:
CT
Sets the value of the counter (CT), 2 bits indicating the continuity of the frame.
Remote command:
13.11.3.2
Triggering on CXPI
Prerequisite: A bus is configured for the CXPI signal to be analyzed.
For the basic trigger settings, proceed in the following way:
1. Press the [Bus] key and select the "Shortcuts" tab.
2. Press "Setup trigger".
3. Tap the "Source" button and select the "Serial bus" trigger source.
4. Select the serial bus that is set to CXPI.
5. Tap "Trigger Type" and select the trigger type to be used for CXPI protocol analy-
sis.
6. Depending on the selected trigger type, more setup conditions can be specified.
CXPI (option R&S