Data and file management
R&S
®
RTO6
461
User Manual 1801.6687.02 ─ 05
If the MSO option R&S
RTO6-B1 is installed, you can save also digital channels and
parallel buses.
Remote command:
Selected sources
Select the channels to be included in data export if "Multi channel" is enabled. Wave-
form1 of up to four input channels can be saved into one file.
Remote command:
Scope
Defines the part of the waveform record that has to be stored.
"Full wave-
form"
Saves the complete waveform record.
"Zoom"
Saves the data included in the zoom area if at least one zoom is
defined for the source waveform. The start and stop values of the
area are shown. If several zooms are defined, select the "Zoom" to be
used for export.
"Cursor"
Saves the data between the cursor lines if at least one cursor mea-
surement is defined for the source waveform. The start and stop val-
ues of the area between the cursor lines are shown. If several cursor
sets are defined, select the "Cursor set" to be used for export.
"Gate"
Saves the data included in the measurement gate if a gated mea-
surement is defined for the source waveform. Select the "Measure-
ment" for which the required gate is defined. The start and stop val-
ues of the gate are shown.
"Manual"
Saves the data between user-defined "Start" and "Stop" values.
Remote command:
Raw (ADC direct)
Enables the export of analog channel data in the raw sample format of the ADC. The
data format is integer 8
bit (signed 8-bit binary format). This format reduces the file size
(1 Byte/sample instead of 4 Bytes/sample in binary files) but decreases the precision of
the values.
If the high definition mode is active, the data format is integer 16
bit, except for peak
detect decimation (8 bit). See
Only y-values are exported, the "Interleaved x/y" option is not available.
Currently, the setting is not available for the export of digital channel data and data of
R&S
RT-ZVC channels.
Save and recall waveform data and results