Remote command:
MEASurement<m>:SPECtrum:ATHReshold
Peak excursion
Defines a relative threshold, the minimum level value by which the waveform must rise
or fall to be considered as a peak. To avoid identifying noise peaks, enter a peak
excursion value that is higher than the noise levels.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
The following figure shows a cursor measurement on a spectrum waveform:
If "Peak excursion" is 30
dB, the peaks 1 to 5 are found. If "Peak excursion" is 20
dB,
also the peaks 6 to 10 are found. The cursor position is on peak 6.
Remote command:
MEASurement<m>:SPECtrum:PEXCursion
8.2
Automatic measurements
The R&S
RTO6 can perform many measurements in parallel. The measurements are
combined in up to 8 measurement groups. In addition, you can run the quick measure-
ment.