Measurements
R&S
®
RTO6
294
User Manual 1801.6687.02 ─ 05
Cu1, Cu2, Cu3, Cu4
The settings for each cursor measurement (or cursor set) are configured on separate
tabs. For each cursor set, a horizontal pair of cursors, a vertical pair of cursors, or both
can be displayed.
Source
Defines the source of the cursor measurement. Any of the input signal, math, reference
or XY waveforms can be selected.
Remote command:
Absolute
Both cursors are set to the absolute peak value.
Remote command:
Next absolute
Cursor 2 is set to the next smaller absolute peak from the current position.
Remote command:
Next left
Cursor 2 is set to the next peak to the left of the current position.
Remote command:
Next right
Cursor 2 is set to the next peak to the right of the current position.
Remote command:
Center frequency
Sets the vertical cursor line Cu1 to the center frequency.
Remote command:
Set center frequency to Cursor
Sets the center frequency to the frequency value that is measured at cursor line Cu1.
Remote command:
Threshold
Defines an absolute threshold as an additional condition for the peak search. Only
peaks that exceed the threshold are detected.
This setting is only available for spectrum waveforms. It is valid for cursor measure-
ments, spectrum measurements and peak search.
Cursor measurements