Remote control commands
R&S
®
RTO6
2569
User Manual 1801.6687.02 ─ 05
Return values:
ValueAtDC
Range:
-1E+9 to 1E+9
Increment: 0.01
*RST:
0
Default unit: dB
Usage:
Query only
DEEMbedding<m>:MATTenuation?
Maximum attenuation of the measurement setup before deembedding, with reference
to a frequency response normalized at DC to 0
dB. The deembedding filter must
amplify the amplitude response by the "Att
max
" value, and thus also increases the noise
by this value. The value also helps to decide whether deembedding is useful for the
defined effective bandwidth.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Return values:
<MaxFilterAtt>
Range:
-1E+9 to 1E+9
Increment: 0.01
*RST:
0
Default unit: dB
Usage:
Query only
DEEMbedding<m>:MGAin?
Maximum gain of the measurement setup before deembedding, with reference to a fre-
quency response normalized at DC to 0
dB. The deembedding filter must attenuate the
amplitude response by this value. The value helps also to avoid an overload of the
ADC.
Suffix:
<m>
.
1..4
Selects the channel that is configured for deembedding.
Return values:
<MaxFilterGn>
Range:
-1E+9 to 1E+9
Increment: 0.01
*RST:
0
Default unit: dB
Usage:
Query only
Deembedding (option R&S
RTO6-K121)