Remote control commands
R&S
®
RTO6
2426
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
The suffix is irrelevant.
Parameters:
<LevelCoupling>
ON | OFF
*RST:
ON
BUS<m>:PARallel:HYSTeresis<n>
<Hysteresis>
Defines the size of the hysteresis for the channel group to avoid the change of signal
states due to noise.
For the parallel bus 1, you can also use
.
Suffix:
<m>
.
1..4
Selects the parallel bus.
<n>
1..4
Selects the channel group:
1 = dig. channels 0..3;
2 = dig. channels 4..7
3 = dig. channels 8..11
4 = dig. channels 12..15
Parameters:
<Hysteresis>
MAXIMUM | MAXimum | ROBUST | ROBust | NORMAL |
NORMal
MAXIMUM = MAXimum
Maximum value that is possible and useful for the signal and its
settings
ROBUST = ROBust
Different hysteresis values for falling and rising edges to avoid
an undefined state of the trigger system.
NORMAL = NORMal
The instrument sets a value suitable for the signal and its set-
tings.
BUS<m>:PARallel:BIT<n>:DESKew
<Deskew>
Sets an individual delay for each digital channel to time-align it with other digital chan-
nels. The deskew value compensates delays that are known from the circuit specifics
or caused by the different length of cables.
For the parallel bus 1, you can also use
.
Suffix:
<m>
.
1..4
Selects the parallel bus.
Mixed signal option (MSO, R&S
RTO6-B1)