Remote control commands
R&S
®
RTO6
2411
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<Mode>
ABS | REL
*RST:
ABS
Usage:
Asynchronous command
EYE<m>:QUALify:GATE:ABSolute:STARt
<Start>
EYE<m>:QUALify:GATE:ABSolute:STOP
<Stop>
Set the absolute time values for gate start and gate end.
Suffix:
<m>
.
1..4
Parameters:
<Start>, <Stop>
Range:
-100E+24 to 100E+24
Increment: 0.01
*RST:
0.01
Example:
Gate definition for eye diagram 2 in absolute time values:
EYE2:QUALify:GATE:COUPling NONE
EYE2:QUALify:GATE:MODE ABS
EYE2:QUALify:GATE:ABSolute:STARt -0.0002
EYE2:QUALify:GATE:ABSolute:STOP 0.0002
Usage:
Asynchronous command
EYE<m>:QUALify:GATE:RELative:STARt
<RelativeStart>
EYE<m>:QUALify:GATE:RELative:STOP
<RelativeStop>
Set the relative values for gate start and gate end.
Suffix:
<m>
.
1..4
Parameters:
<RelativeStart>,
<RelativeStop>
Range:
0 to 100
Increment: 0.1
*RST:
100
Default unit: %
Example:
Gate definition for eye diagram 2 in relative values:
EYE2:QUALify:GATE:COUPling NONE
EYE2:QUALify:GATE:MODE REL
EYE2:QUALify:GATE:RELative:STARt 10
EYE2:QUALify:GATE:RElative:STOP 90
Usage:
Asynchronous command
Protocols