Remote control commands
R&S
®
RTO6
2275
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataLaneCount>
Range:
1 to 4
Increment: 1
*RST:
4
BUS<m>:MPHY:DZERo:SOURce
<DataSource>
Sets the source for the lane.
Suffix:
<m>
.
1..4
Parameters:
<DataSource>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST:
C1W1
BUS<m>:MPHY:DONE:SOURce
<DataSource>
BUS<m>:MPHY:DTWO:SOURce
<DataSource>
BUS<m>:MPHY:DTHRee:SOURce
<DataSource>
Selects the source of the corresponding data lane.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<DataSource>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | NONE
*RST:
DONe: C2W1, DTWo: C3W1, DTHree: C4W1
BUS<m>:MPHY:DZERo:THReshold
<ThresholdD0>
BUS<m>:MPHY:DONE:THReshold
<ThresholdD1>
BUS<m>:MPHY:DTWO:THReshold
<ThresholdD2>
BUS<m>:MPHY:DTHRee:THReshold
<ThresholdD3>
Sets the threshold value for the respective lane.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ThresholdD1>
<ThresholdD2>
<ThresholdD3>
Range:
0 to 1
Increment: 0.01
*RST:
0.1
Default unit: V
Protocols