Remote control commands
R&S
®
RTO6
2247
User Manual 1801.6687.02 ─ 05
Parameters:
<DataSource>
NONE | C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 |
M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4
*RST:
DNZero: NONE, DPONe: C2W1, DPTWo: C3W1,
DPTHree: C4W1
BUS<m>:DPHY:DNZero:PROBe
<ProbeDN0>
BUS<m>:DPHY:DPZero:PROBe
<ProbeDP0>
BUS<m>:DPHY:DPONe:PROBe
<ProbeDP1>
BUS<m>:DPHY:DPTWo:PROBe
<ProbeDP2>
BUS<m>:DPHY:DPTHree:PROBe
<ProbeDP3>
Selects the type of probe used for the corresponding data lane.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<ProbeDP3>
DIFFerential | SINGle
*RST:
SINGle
BUS<m>:DPHY:CP:HSPeed:HYSTeresis
<HysteresisHSCP>
BUS<m>:DPHY:DPZero:HSPeed:HYSTeresis
<HysteresisHSDP0>
BUS<m>:DPHY:DPONe:HSPeed:HYSTeresis
<HysteresisHSDP1>
BUS<m>:DPHY:DPTWo:HSPeed:HYSTeresis
<HysteresisHSDP2>
BUS<m>:DPHY:DPTHree:HSPeed:HYSTeresis
<HysteresisHSDP3>
Sets a value for the hysteresis of the respective lane.
Suffix:
<m>
.
1..4
Selects the serial bus.
Parameters:
<HysteresisHSDP3> Range:
0 to 0.5
Increment: 0.01
*RST:
0.2
Default unit: V
BUS<m>:DPHY:CP:HSPeed:THReshold
<ThresholdHSCP>
BUS<m>:DPHY:DPZero:HSPeed:THReshold
<ThresholdHSDP0>
BUS<m>:DPHY:DPONe:HSPeed:THReshold
<ThresholdHSDP1>
BUS<m>:DPHY:DPTWo:HSPeed:THReshold
<ThresholdHSDP2>
BUS<m>:DPHY:DPTHree:HSPeed:THReshold
<ThresholdHSDP3>
Sets the threshold value for the digitization of the respective high speed data line.
Suffix:
<m>
.
1..4
Selects the serial bus.
Protocols