Remote control commands
R&S
®
RTO6
2018
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Parameters:
<Polarity>
NORMal | INVert
NORMal
Normal (non-inverted) data signal polarity
INVert
Inverted data signal polarity
*RST:
NORMal
BUS<m>:ETHernet:THReshold:HIGH
<ThresholdUpper>
Sets the upper threshold value for the signal digitization. If the signal value is higher
than the this threshold, the signal state is considered high.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdUpper>
Range:
0 to 10
Increment: 0.01
*RST:
0.5
Default unit: V
BUS<m>:ETHernet:THReshold:LOW
<ThresholdLower>
Sets the lower threshold value for the signal digitization. If the signal value is below this
threshold, the signal state is considered low.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdLower>
Range:
-10 to 0
Increment: 0.01
*RST:
-0.5
Default unit: V
BUS<m>:ETHernet:PRESet
<ThresholdPreset>
Sets the thresholds to predefined or individually definable voltage levels.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdPreset>
T0 | T100 | TX0 | TX100 | MANual
T0
Sets the thresholds to the default values for 10BASE-T
(0
meters): upper threshold to 1.25
V, lower threshold to -1.25
V
Protocols