Remote control commands
R&S
®
RTO6
1780
User Manual 1801.6687.02 ─ 05
LENGth
The length error indicates that the frame ended prematurely, or
an array in the frame had too few elements. The amount of bits
that the software expected (based upon the user's frame
description) was not found before the frame was terminated.
This might occur because a new frame synchronized, or a gap
appeared between the bits.
UNKNown
Unknown error
INComplete
The frame ended prematurely because it extends past the end
of the record.
CRC
Checksum error in cyclic redundancy check (error in data)
PARity
Parity bit error, indicating a transmission error (only available if a
parity is configured)
*RST:
OK
Usage:
Query only
BUS<m>:CMSB:RESult<n>:STARt?
Returns the start time of the frame.
Suffix:
<m>
.
1..4
<n>
*
Return values:
<Start>
Range:
-100E+24 to 100E+24
Increment: 100E-12
*RST:
0
Default unit: s
Usage:
Query only
BUS<m>:CMSB:RESult<n>:STOP?
Returns the stop time of the frame.
Suffix:
<m>
.
1..4
<n>
*
Return values:
<Stop>
Range:
-100E+24 to 100E+24
Increment: 100E-12
*RST:
0
Default unit: s
Protocols