![R&S RTO6 User Manual Download Page 1762](http://html2.mh-extra.com/html/rands/rto6/rto6_user-manual_30646461762.webp)
Remote control commands
R&S
®
RTO6
1762
User Manual 1801.6687.02 ─ 05
BUS<m>:CMSB:BITRate:ENABle
<BitrateEnable>
Enables the bit rate settings (must always be enabled for the coding standard "NRZ
Unclocked", and also for triggering on signals in any coding standard).
Suffix:
<m>
.
1..4
Parameters:
<BitrateEnable>
ON | OFF
ON
Bit rate settings enabled
OFF
Bit rate settings disabled
*RST:
OFF
BUS<m>:CMSB:BITRate:VALue
<Bitrate>
Sets the transmission speed setting for the custom serial bus data signal.
Suffix:
<m>
.
1..4
Parameters:
<Bitrate>
Range:
300 to 2E9
Increment: 100E3
*RST:
10E6
Default unit: bps
BUS<m>:CMSB:GAPTime:ENABle
<GapTimeEnable>
Enables the gap time settings (must always be enabled for the coding standard "NRZ
Unclocked", and also for triggering on signals in any coding standard).
Suffix:
<m>
.
1..4
Parameters:
<GapTimeEnable>
ON | OFF
ON
Gap time settings enabled
OFF
Gap time settings disabled
*RST:
OFF
BUS<m>:CMSB:GAPTime:VALue
<MinGapTime>
Sets a minimum gap time for synchronization.
Suffix:
<m>
.
1..4
Protocols