Remote control commands
R&S
®
RTO6
1758
User Manual 1801.6687.02 ─ 05
BUS<m>:CMSB:NRZ:CPHase
<ClockPhase>
Selects the phase of the custom serial bus clock signal for the coding standard "NRZ
Clocked", depending on
Suffix:
<m>
.
1..4
Parameters:
<ClockPhase>
FEDGe | SEDGe
FEDGe
Sets the clocking transaction to be on the first edge:
If Clock Polarity = "IDLLow", data are captured on the clock's ris-
ing edge (low-to-high transition) and propagated on a falling
edge
If Clock Polarity = "IDLHigh", data are captured on the clock's
falling edge (high-to-low transition) and propagated on a rising
edge
SEDGe
Sets the clocking transaction to be on the second edge:
If Clock Polarity = "IDLLow", data are captured on the clock's
falling edge (high-to-low transition) and propagated on a rising
edge
If Clock Polarity = "IDLHigh", data are captured on the clock's
rising edge (low-to-high transition) and propagated on a falling
edge
*RST:
FEDGe
BUS<m>:CMSB:NRZ:ENBLe
<SourceNRZEnable>
Selects the input source for the custom serial bus enable signal.
If an input is chosen, signals will be only decoded while this channel is in the enabled
state. This allows you to mark a time when the signal on the selected source is active
and when not.
For triggering on a serial bus, analog channels "C1"–"C4" are required. Otherwise, if no
serial bus trigger has been selected, permitted source selections include the mathe-
matical channels and the reference channels.
Suffix:
<m>
.
1..4
Parameters:
<SourceNRZEnable> C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4 | D0 | D1 | D2 | D3 | D4 | D5 | D6 |
D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15
Protocols