Remote control commands
R&S
®
RTO6
1754
User Manual 1801.6687.02 ─ 05
BUS<m>:CMSB:MANChester:DATA
<SourceMANData>
Selects the source channel for the data signal. For triggering on a serial bus, analog
channels "C1"
–
"C4" are required. Otherwise, if no serial bus trigger has been selected,
permitted source selections include the mathematical channels and the reference
channels .
Suffix:
<m>
.
1..4
Parameters:
<SourceMANData>
C1W1 | C2W1 | C3W1 | C4W1 | M1 | M2 | M3 | M4 | M5 | M6 |
M7 | M8 | R1 | R2 | R3 | R4
*RST:
C1W1
BUS<m>:CMSB:MANChester:POLarity
<PolarityData>
Selects the polarity of the custom serial bus data signal in Manchester coding stand-
ards.
Suffix:
<m>
.
1..4
Parameters:
<PolarityData>
NORMal | INVert
NORMal
Manchester or Manchester II polarity remains unchanged
INVert
Manchester polarity is inverted and becomes Manchester II
polarity, Manchester II polarity is inverted and becomes Man-
chester polarity
*RST:
NORMal
BUS<m>:CMSB:MANChester:THReshold:HIGH
<ThresholdUpper>
Sets the upper threshold for data signal digitization in Manchester coding standards.
Suffix:
<m>
.
1..4
Parameters:
<ThresholdUpper>
Range:
-25 to 25
Increment: 0.1
*RST:
5
Default unit: V
BUS<m>:CMSB:MANChester:THReshold:LOW
<ThresholdLower>
Sets the lower threshold for data signal digitization in Manchester coding standards.
Protocols