Remote control commands
R&S
®
RTO6
1588
User Manual 1801.6687.02 ─ 05
Setting parameters:
<MaskTestName>
String with the name of the mask test
<MaskSegmIdx>
Number of the segment. Counting starts from 0.
Usage:
Setting only
Asynchronous command
MTESt:SEGMent:REMove
<MaskTestName>, <MaskSegmIdx>
Removes the specified segment from the mask definition.
Setting parameters:
<MaskTestName>
String with the name of the mask test
<MaskSegmIdx>
Number of the segment. Counting starts from 0.
Usage:
Setting only
Asynchronous command
MTESt:SEGMent:CLEar
<MaskTestName>
Deletes all mask segments of the indicated mask test.
Setting parameters:
<MaskTestName>
String with the name of the mask test
Usage:
Setting only
Asynchronous command
MTESt:SEGMent:REGion
<MaskTestName>, <MaskSegmIdx>,<Region>
MTESt:SEGMent:REGion?
<MaskTestName>, <MaskSegmIdx>
Defines the region of the segment that builds the mask.
Parameters:
<Region>
UPPer | LOWer | INNer
UPPer
the segment points are connected to a line, the display area
above this line is the mask segment
LOWer
the segment points are connected to a line, the display area
below this line is the mask segment
INNer
the segment points form a closed geometrical shape, which is
the mask segment
*RST:
INNer
Parameters for setting and query:
<MaskTestName>
String with the name of the mask test
<MaskSegmIdx>
Number of the segment. Counting starts from 0.
Mask testing