Remote control commands
R&S
®
RTO6
1573
User Manual 1801.6687.02 ─ 05
TIMebase:RACTime?
Queries the required acquisition time. If FFT gating is used and the resolution BW is
set to constant, record length can be extended to acquire the required number of sam-
ples. In this case, the required acquisition time differs from the adjusted acquisition
time (
Return values:
<RqrdAcqTime>
Required acquisition time for FFT
Range:
125E-12 to 100E+3
*RST:
0.5
Default unit: s
Usage:
Query only
Asynchronous command
CALCulate:MATH<m>:FFT:GATE:ABSolute:STARt
<Start>
Defines the starting value for the gate.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<Start>
Range:
-100E+24 to 100E+24
Increment: 0.01
*RST:
0
Usage:
Asynchronous command
CALCulate:MATH<m>:FFT:GATE:ABSolute:STOP
<Stop>
Defines the end value for the gate.
Suffix:
<m>
.
1..8
math waveform
Parameters:
<Stop>
Range:
-100E+24 to 100E+24
Increment: 0.01
*RST:
0.01
Usage:
Asynchronous command
CALCulate:MATH<m>:FFT:GATE:MODE
<Mode>
Defines whether the gate settings are configured using absolute or relative values.
Suffix:
<m>
.
1..8
math waveform
Spectrum analysis