Remote control commands
R&S
®
RTO6
1475
User Manual 1801.6687.02 ─ 05
CURSor<m>:X2ENvelope
<EnvelopeCurve2>
If the waveform arithmetics are set to envelope curve (see
is set to "ON", this setting defines how the second horizontal cursor is posi-
tioned.
Suffix:
<m>
.
1..*
Selects the cursor set. 4 cursor sets are available.
Parameters:
<EnvelopeCurve2>
MIN | MAX
MIN
The horizontal cursor is set to the crossing point of the vertical
cursor with the minimum waveform envelope.
MAX
The horizontal cursor is set to the crossing point of the vertical
cursor with the maximum waveform envelope.
*RST:
MAX
Usage:
Asynchronous command
CURSor<m>:LABel[:LABel]
<ShowLabel>
Shows the cursor labels in the diagram.
Suffix:
<m>
.
1..4
Selects the cursor set.
Parameters:
<ShowLabel>
ON | OFF
*RST:
ON
Usage:
Asynchronous command
CURSor<m>:LABel:X<1..2>
<Name>
Defines a label to be displayed with the vertical cursors. X1 is the left cursor line, and
X2 the right one.By default, the cursors are labeled as Cu1.X1, Cu1.X2, Cu2.X1,
Cu2.X2, Cu3.X1, Cu3.X2, Cu4.X1, Cu4.X2.
Suffix:
<m>
.
1..4
Selects the cursor set.
Parameters:
<Name>
String with the cursor label
Usage:
Asynchronous command
Cursor measurements