Remote control commands
R&S
®
RTO6
1451
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<ScaleFactor>
Range:
-1E+6 to 1E+6
Increment: 0.1
*RST:
1
Usage:
Asynchronous command
REFCurve<m>:RESCale:VERTical:OFFSet
<Offset>
Moves the reference waveform vertically. Like vertical offset of channel waveforms, the
offset of a reference waveform is subtracted from the measured value.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<Offset>
Negative values shift the waveform up, positive values shift it
down.
Range:
-100E+24 to 100E+24
Increment: 1E-6
*RST:
0
Default unit: V
Usage:
Asynchronous command
REFCurve<m>:HMODe
<HorizontalMode>
Selects the coupling of horizontal settings.
Suffix:
<m>
.
1..4
Reference waveform
Parameters:
<HorizontalMode>
ORIGinal | COUPled
ORIGinal
Horizontal scaling and reference point of the source waveform
are used.
COUPled
The current horizontal settings of the diagram are used.
*RST:
ORIGinal
Usage:
Asynchronous command
Waveform analysis