Remote control commands
R&S
®
RTO6
1436
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
TRIGger<m>:ANEDge:GND
<Ground>
Connects the analog signal to the ground.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Ground>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
TRIGger<m>:ANEDge:SLOPe
<Slope>
Sets the edge for the trigger event.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Slope>
POSitive | NEGative
See
Chapter 24.4.3, "Slope parameter"
*RST:
POSitive
Usage:
Asynchronous command
TRIGger<m>:EXTern:OVERload
<Overload>
The query returns the overload status of the external trigger input.
:TRIGger:EXTern:OVERload 0
confirms the information in the message box, it has
same effect as OK.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Overload>
ON | OFF
ON | 1
Iindicates an overload of the probe
OFF | 0
In a query: no overlaod.
Ssetting: confirms the information in the message box
*RST:
OFF
Usage:
Asynchronous command
Trigger