Remote control commands
R&S
®
RTO6
1434
User Manual 1801.6687.02 ─ 05
Usage:
Asynchronous command
24.9.22
External trigger input
External trigger signals, which are connected to the external trigger input, can be trig-
gered with an edge trigger.
TRIGger<m>:ANEDge:COUPling
<Coupling>
Sets the coupling for the external trigger signal.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Coupling>
DC | DCLimit | AC
DC
Direct connection with 50
Ω
termination, passes both DC and
AC components of the trigger signal.
DCLimit
Direct connection with 1 M
Ω
termination, passes both DC and
AC components of the trigger signal.
AC
Connection through DC capacitor, removes DC and very low-fre-
quency components.
Usage:
Asynchronous command
TRIGger<m>:ANEDge:FILTer
<Filter>
Sets a filter for the external trigger signal to reject high or low frequencies.
Suffix:
<m>
.
1..3
Only 1 = A-trigger, 2 | 3 = not available. Can be omitted.
Parameters:
<Filter>
OFF | LFReject | RFReject
OFF
The trigger signal is not filtered.
LFReject
Frequencies higher than the "Cut-off" frequency are rejected,
lower frequencies pass the filter.
You can adjust the "Cut-off" frequency using the
kHz.
Trigger