Remote control commands
R&S
®
RTO6
1372
User Manual 1801.6687.02 ─ 05
ZVC:Z<m>:I<n>:SHUNt:MAXVoltage
<ExtShuntMeasRg>
Sets the maximum voltage for the external shunt.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<ExtShuntMeasRg> V450m | V045m
V450m
: 450 mV
V045m
: 45 mV
*RST:
V450m
Usage:
Asynchronous command
ZVC:Z<m>:I<n>:SKEW
<DeskewOffset>
Sets the skew offset value for the current probe channel. This is a delay value, that is
known from the circuit specifics but cannot be compensated by the instrument auto-
matically.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<DeskewOffset>
Range:
-6E-6 to 6E-6
Increment: 200E-9
*RST:
0
Default unit: s
Usage:
Asynchronous command
ZVC:Z<m>:I<n>[:STATe]
<State>
Enables the corresponding current channel of the probe.
Suffix:
<m>
.
1..2
<n>
1..4
Parameters:
<State>
ON | OFF
*RST:
OFF
Usage:
Asynchronous command
Acquisition and setup