Remote control commands
R&S
®
RTO6
1341
User Manual 1801.6687.02 ─ 05
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Impedance>
Range:
0.1 to 100E+3
Increment: 1
*RST:
50
Default unit: Ω
Usage:
Asynchronous command
CHANnel<m>:OVERload
<Overload>
Retrieves the overload status of the specified channel from the status bit. When the
overload problem is solved, the command resets the status bit.
The overload status is returned asynchronously. Therefore, it is not possible to assign
an overload to a specific acquisition during continuous acquisition. This assignment is
only possible for a single acquisition.
Suffix:
<m>
.
1..4
Selects the input channel.
Parameters:
<Overload>
ON | OFF
Use OFF to reset the overload status bit.
*RST:
OFF
Example:
ACQuire:COUNt 1
SINGle
CHANnel2:OVERload?
Queries the overload status of channel 2 after a single acquisi-
tion.
CHANnel2:OVERload OFF
Resets the overload status bit.
Usage:
Asynchronous command
CHANnel<m>:FILDesign
<FilterDesignTyp>
Selects the type of filter to be used.
Suffix:
<m>
.
1..4
Parameters:
<FilterDesignTyp>
WALL | GAUSsian
WALL
: Brick wall filter
*RST:
WALL
Usage:
Asynchronous command
Acquisition and setup