Deembedding (option R&S
RTO6-K121)
R&S
®
RTO6
1213
User Manual 1801.6687.02 ─ 05
Normalize
Starts the normalization of the setup without the probe or cable to be measured. The
frequency correction and deskew are determined and applied in the next measurement
step.
If normalization was successful, it is indicated by an icon, and you can display the
"Normalization Result" - the normalized magnitude response.
Save normalization setup
Saves the results of the normalization step to an XML file.
20.4.1.3
Measurement
Open [App Cockpit]> "Analysis" tab > "Proven Probe / Cable" > "Normalize setup">
"Next"
1. To characterize a cable, connect it to the cable coming from the
connector of
the pulse source and the "Device" input channel. Use an adapter if necessary.
Figure 20-2: Cable measurement
2. To characterize a probe, you need the R&S
RT-ZF30 probe test fixture.
Proven probe, proven cable