Power analysis (option R&S
RTO6-K31)
R&S
®
RTO6
1174
User Manual 1801.6687.02 ─ 05
Trigger level on ← Turn on
Triggers the beginning of the measurements at the moment the AC or DC input voltage
reaches the set value.
Remote command:
Turn off
Enables the configuration of the turn off time measurement.
Steady state level- Turn off ← Turn off
Sets the percentage of the steady state level of the DC output that has to be reached.
Remote command:
Trigger level ← Turn off
Triggers the beginning of the measurements at the moment the AC or DC input voltage
reaches the set value.
Remote command:
Time ← Turn off
Sets the time, the start of the measurement of the turn off time is delayed with, after
the trigger point.
Remote command:
Report comment
In this field you can write a comment that is displayed in the measurement report, for
example, a measurement description.
Execute
Starts the " Power Quality" measurement.
Remote command:
19.5.3
Switching loss
The "Switching Loss" analysis measures the power and energy losses of a switching
device, that occur during the switching phases and the conduction phase of the switch-
ing transistor.
Required probes:
●
Differential voltage probe
●
Current probe
Power path