Operating the Instrument
R&S
®
RTO2000
84
Getting Started 1332.9719.02 ─ 10
, the signal icons Ch1Wfm1 and Ch2Wfm1 show the main settings,
and the waveforms are displayed in diagrams. All other waveforms are minimized
and shown in the signal icon.
Figure 7-3: Signal label on the sidebar
1 = Vertical scale
2 = Vertical position
3 = Coupling and Termination
4 = Acquisition mode
5 = Clipping indication is shown if the signal is clipped at the top and/or bottom.
6 = Offset
7 = Actual bandwidth, can be less than the instrument bandwidth depending on the number of
active channels and other settings
If the sidebar contains many icons and not all icons are visible, touch one of the
icons and move it up or down until the required icon appears.
You can also switch off the sidebar: "Display" menu > "Diagram Layout" > "Show
signal bar" .
Trigger position and trigger level (4, 5)
The blue markers show the horizontal position of the trigger and the vertical trig-
ger level. You can touch and move the trigger markers in the diagram to set the
positions. The trigger point is the zero point of the diagram.
The trigger position can be moved outside the diagram. A red trigger position
marker indicates that the trigger position is not visible.
Reference point (6)
The reference point marks the rescaling center. If you modify the time scale, the
reference point remains fixed on the screen, and the scale is stretched or com-
pressed to both sides of the reference point.
You can define the position of the reference point (Horizontal), and its time dis-
tance from the trigger point of the diagram (Position).
Touchscreen Display