Trying out the instrument
R&S
®
RTE
48
Getting Started 1326.1026.02 ─ 10
The dialog box turns into a mini box that contains only the "Wfm Arithmetic"
setting.
5. Select
Envelope
and check the waveform.
The instrument sets the "Decimation" type automatically to
Peak detect
to dis-
play the correct envelope waveform.
6. Select
Average
and check the waveform.
7. Tap the "Maximize" icon in the mini dialog box.
The complete "Acquisition" dialog box is restored.
8. Close the dialog box by tapping .
6.3
Changing the waveform scaling and position
As you can see on the y-axis of the display, the calibration signal has a vertical
offset of about 500
mV. The value can differ.
This value is also indicated in the signal icon for channel1 (3rd row). If you press
the [Ch1] key, the "Vertical" settings dialog box also displays the "Offset" value.
The offset is the DC component of the signal.
Changing the waveform scaling and position