CHANnel<m>:OFFSet
<Offset>
The offset voltage is subtracted to correct an offset-affected signal.
Suffix:
<m>
.
Selects the input channel (1, 2).
Parameters:
<Offset>
Offset value in V
Range:
Values depend on vertical scale and probe attenua-
tion.
*RST:
0
CHANnel<m>:BANDwidth
<BandwidthLimit>
Selects the bandwidth limit for the indicated channel.
Suffix:
<m>
.
Selects the input channel (1, 2).
Parameters:
<BandwidthLimit>
FULL | B20
FULL
Use full bandwidth.
B20
Bandwidth limit 20 MHz.
*RST:
FULL
CHANnel<m>:POLarity
<Polarity>
Turns the inversion of the signal amplitude on or off. To invert means to reflect the volt-
age values of all signal components against the ground level. Inversion affects only the
display of the signal but not the trigger.
Suffix:
<m>
.
Selects the input channel (1, 2).
Parameters:
<Polarity>
NORMal | INVerted
*RST:
NORM
CHANnel<m>:OVERload
<Overload>
Retrieves the overload status of the specified channel from the status bit. When the
overload problem is solved, the command resets the status bit.
Suffix:
<m>
.
Selects the input channel (1, 2).
Acquisition and Setup