9 Mixed Signal Operation (Option R&S
RTC-
B1)
The R&S
RTC1000 has a connector for the R&S
RT-ZL03 logic probe to add 8 digital
logic inputs. To use the logic analysis, you need the option R&S
RTC-B1. The logic
probe comes with the option.
Figure 9-1: Optional logic probe R&S
RT-ZL03
Ensuring accurate measurement results
The logic analyzer option R&S
RTC-B1 with connected probe leads is considered as a
test probe, according to EN 61326-2-1, clause 5.2.4.101. Therefore, the measure-
ments are sensitive to electromagnetic interference. Consider additional shielding
methods to avoid interference.
To ensure accurate measurements, attach the ground lead from the pod to the ground
of the device under test if any logic channel of the pod is used for data capture. The
ground lead improves signal fidelity to the oscilloscope.
9.1 Using Logic Channels
You can display and adjust logic channels in various ways. Make sure to set the
threshold to identify the logic states (high and low) .
●
"To set the threshold for logic states"
●
"To activate and scale logic channels"
●
"To reset the position and size of logic channels"
Using Logic Channels