![R&S RTA4000 User Manual Download Page 645](http://html2.mh-extra.com/html/rands/rta4000/rta4000_user-manual_3161987645.webp)
Remote control commands
R&S
®
RTA4000
645
User Manual 1335.7898.02 ─ 08
Suffix:
<b>
.
1..4
<n>
*
Selects the frame (1...n).
Return values:
<FrameType>
DATA | REMote | ERR | OVLD
Data, remote, error or overload frame
Usage:
Query only
BUS<b>:CAN:FRAMe<n>:STATus?
Returns the overall state of the specified frame.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame (1...n).
Return values:
<FrameStatus>
OK | BTST | CRCD | ACKD | CRC | EOFD | NOACk |
INSufficient
OK: frame is valid.
BTST: bit stuffing error occured
CRCD: wrong CRC delimiter occured
ACKD: Wrong ACK delimiter occured
CRC: cyclic redundancy check failed
EOFD: wrong end of frame
NOACk: acknowlegde is missing
INSufficient: frame is not completely contained in the acquisition.
The acquired part of the frame is valid.
Usage:
Query only
BUS<b>:CAN:FRAMe<n>:STARt?
Returns the start time of the specified frame.
Suffix:
<b>
.
1..4
<n>
*
Selects the frame (1...n).
Return values:
<StartTime>
Default unit: s
Usage:
Query only
Serial bus analysis