Measurement Basics
R&S
®
NRX
47
User Manual 1178.5566.02 ─ 07
5.5
Settings Conflict
A settings conflict can occur for the following reasons:
●
The sensor assigned to the measurement does not support a set value.
If it is a numeric value, the suitable range for the sensor is given in the tooltip.
●
The sensor assigned to the measurement does not support the measurement type.
●
Other contradictory settings, for example the lower limit value is higher than the
upper limit value.
Contradictory settings are allowed so that you are not hampered in your workflow. But
they cause an error message in the notification center. Furthermore, the contradictory
setting is highlighted and the control elements in the hierarchies above that are leading
to this setting are highlighted, too. Thus, you can follow the problem across the hierar-
chies to solve the settings conflict. The only control element that is not highlighted due
to a settings conflict is the measurement value. The measurement value is only high-
lighted in red when it is violating the set limits, as shown in
Example: The sensor does not support the measurement type
The notification center indicates an error, but the no control element is highlighted.
The error message reports a settings conflict.
Solving the error
1. Tap the
displayed measurement value or graphic
.
The "Measurement Settings" dialog is displayed.
The "Measurement Type" is highlighted.