Note:
Even if the exclusion interval exceeds the timeslot because, for example, its right
limit is outside the timeslot, correct results are obtained. In the extreme case, where
the interval length has been set to a value greater than the timeslot length, 0 W is out-
put as the measured power. No error message is output.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
Range:
0.0 to 0.1
*RST:
0.0
Default unit: s
Manual operation:
See
CALCulate<Measurement>[:POWer]:TSLot[:AVG][:EXCLude]:MID[:STATe]
<value>
Enables or disables the blanking out of time intervals in the timeslots.
Suffix:
<Measurement>
.
1 to 4
Measurement channel
Parameters:
<value>
*RST:
OFF
Manual operation:
See
14.4.10
Statistics
Further information:
●
14.4.10.1
Statistics Measurement Settings
In a statistics measurement, you can use commands that combine several setting com-
mands. They are described in
Chapter 14.4.10.2, "Combining Statistics Commands"
The same principle is used for the calculation functions, see
For time gate settings, see also
[SENSe<Sensor>:]STATistics:AVERage?
......................................................................... 251
[SENSe<Sensor>:]STATistics:OFFSet[:TIME]
................................................................... 251
[SENSe<Sensor>:]STATistics:PEAK?
...............................................................................251
CALCulate<Measurement>:STATistics:POWer:AVG:DATA?
............................................... 251
CALCulate<Measurement>:STATistics:APERture
..............................................................252
Measurement Settings and Results