Remote Control Commands
R&S
®
NRPxxS(N)
90
User Manual 1177.5079.02 ─ 10
"POWer:BURSt:AVG"
Burst
In remote control, this measurement mode is very similar to the
ContAv mode. The integration time is, however, not predefined
but determined by the sensor with the aid of a burst detector.
The start of a burst is detected when the measurement signal
rises above the set trigger level. The end is set when the signal
drops below the trigger threshold.
[SENSe<Sensor>:][POWer:]BURSt:DTOLerance
the time interval during which the signal level must be below the
trigger level so that the end of the burst can be detected. In the
Burst mode, the set trigger source is ignored and
INT
is implicitly assumed.
"POWer:TSLot:AVG"
Timeslot
The power is simultaneously measured in up to 32 time win-
dows.
The number of time windows is set with
The length of a time window is determined via
[SENSe<Sensor>:][POWer:]TSLot[:AVG]:WIDTh
measurement result is represented by a vector that can contain
up to 32 indices and contains the power of a time window at
each index.
"XTIMe:POWer"
Trace
In this mode, power over time is measured. Therefore a number
of measurement points are defined (
) where the length of an individual measure-
ment is determined from the ratio of total time
(
measurement points.
*RST:
"POWer:AVG"
FETCh<Sensor>:ARRay[:POWer][:AVG]?
Queries the last valid measurement result of a measurement with enabled data buffer
mode.
Usage:
Query only
FETCh<Sensor>[:SCALar][:POWer][:AVG]?
Queries the last valid measurement result.
Usage:
Query only
Configuring and Retrieving Results