Error Messages and Troubleshooting
R&S
®
NRP2
336
User Manual 1173.9157.02 ─ 03
8.2 SCPI-Specific Error Messages
The list below contains error messages that may occur in the instrument. Negative error
numbers are defined in the SCPI standard, positive error numbers identify device-specific
errors.
The left column of the following table gives the error code. In the right column, the text of
the error message displayed or entered in the error/event queue is printed in bold. An
additional explanation is given below this text.
8.2.1 No error
Error code
"Error text returned upon queue query"
Explanations
0
"No error"
This message is displayed if the error queue is empty.
8.2.2 Command Error
Command errors; causes bit 5 in the ESR to be set.
Error code
Error text returned upon queue query
Explanations
-100
"Command error"
Command faulty or invalid.
-101
"Invalid character"
The command contains a character which is invalid for that type.
-102
"Syntax error"
The command is invalid.
-103
"Invalid separator"
Command contains an illegal character where a separator is expected.
-104
"Data type error"
Command contains an invalid data element.
-108
"Parameter not allowed"
The command contains too many parameters.
-109
"Missing parameter"
The command contains fewer parameters than required.
-111
"Header separator error"
The header contains an illegal delimiter.
-112
"Program mnemonic too long"
The header contains more than 12 characters.
SCPI-Specific Error Messages
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...