Remote Control - Commands
R&S
®
NRP2
309
User Manual 1173.9157.02 ─ 03
exceeded since the last readout of the Operation Upper Limit Fail Status Register. Details
of the behavior are defined by the transition register. The limit value can be set with the
command
CALC:LIM:UPP:DATA <float_value>
. Command
CALC:LIM:UPP:STAT ON
configures the PTRansition and NTRansition registers so that
the corresponding bit is set in the EVENt register when the displayed value exceeds the
upper limit value.
The Operation Upper Limit Fail Status Register can be read by the commands:
STATus:OPERation:ULIMit[:SUMMary]:CONDition?
STATus:OPERation:ULIMit[:SUMMary][:EVENt]?
.
Table 6-28: Meaning of bits used in the Operation Upper Limit Fail Status Register
Bit No.
Meaning
0
Not used
1
Window 1 Upper Limit Fail
The measured value returned by the first Calculate block exceeds the upper limit value.
2
Window 2 Upper Limit Fail
The measured value returned by the second Calculate block exceeds the upper limit value.
3
Window 3 Upper Limit Fail
The measured value returned by the third Calculate block exceeds the upper limit value.
4
Window 4 Upper Limit Fail
The measured value returned by the fourth Calculate block exceeds the upper limit value.
5
Window 5 Upper Limit Fail
The measured value returned by the fifth Calculate block exceeds the upper limit value.
6
Window 6 Upper Limit Fail
The measured value supplied by the sixth Calculate block exceeds the upper limit value.
7
Window 7 Upper Limit Fail
The measured value supplied by the seventh Calculate block exceeds the upper limit value.
8
Window 8 Upper Limit Fail
The measured value supplied by the eighth Calculate block exceeds the upper limit value.
9 to14
Not used
15
Bit 15 will never be used.
6.13.17 Questionable Power Status Register
The CONDition register contains information as to whether the measured power values
are questionable.
The Questionable Power Status Register can be read by the following commands:
STATus:QUEStionable:POWer[:SUMMary]:CONDition?
STATus:QUEStionable:POWer[:SUMMary][:EVENt]?
STATus
Summary of Contents for NRP2
Page 1: ...R S NRP2 Power Meter User Manual User Manual 2 1173 9157 02 03 Test Measurement...
Page 20: ......
Page 24: ...Contents R S NRP2 6 User Manual 1173 9157 02 03...
Page 40: ...Putting into Operation R S NRP2 22 User Manual 1173 9157 02 03 Starting the R S NRP2...
Page 352: ...Maintenance R S NRP2 334 User Manual 1173 9157 02 03 Storage...
Page 366: ...Error Messages and Troubleshooting R S NRP2 348 User Manual 1173 9157 02 03 Bootmenu...
Page 382: ...Remote control interfaces R S NRP2 364 User Manual 1173 9157 02 03 USB interface...
Page 394: ...Standard Configurations R S NRP2 376 User Manual 1173 9157 02 03...
Page 422: ...Compatibility Information R S NRP2 404 User Manual 1173 9157 02 03...