![R&S NRP Series User Manual Download Page 119](http://html1.mh-extra.com/html/rands/nrp-series/nrp-series_user-manual_774882119.webp)
Remote control commands
R&S
®
NRPxxA(N)
119
User Manual 1177.6017.02 ─ 08
SYSTem:COMMunicate:NETWork[:COMMon]:HOSTname
<hostname>
Effective only for the R&S
NRP LAN power sensors.
Sets the individual hostname of the power sensor.
In a LAN that uses a DNS server (domain name system server), you can access each
connected instrument using a unique hostname instead of its IP address. The DNS
server translates the hostname to the IP address. Using a hostname is especially use-
ful if a DHCP server is used, as a new IP address can be assigned each time the
instrument is restarted.
The power sensor performs the change of the hostname immediately after the com-
mand is sent. For this purpose, the power sensor restarts its connection to the network,
which can take several seconds. During this time, you cannot address the power sen-
sor. After the restart, you can only address the power sensor using the newly set host-
name.
Note:
It is recommended that you do not change the default hostname to avoid prob-
lems with the network connection. However, if you change the hostname, be sure to
use a unique name.
Parameters:
<hostname>
Example:
SYST:COMM:NETW:COMM:HOST
'powersensor-2nd-floor'
Sets
powersensor-2nd-floor
as new hostname.
SYSTem:COMMunicate:NETWork:IPADdress
<ipaddress>
Effective only:
●
For the R&S
NRP LAN power sensors.
●
SYSTem:COMMunicate:NETWork:IPADdress:MODE
is set to
STATic
.
Sets the IP address of the sensor.
Parameters:
<ipaddress>
Example:
SYST:COMM:NETW:IPAD '192.168.10.29'
Sets
192.168.10.29
as IP address.
Manual operation:
See
SYSTem:COMMunicate:NETWork:IPADdress:GATeway
<gateway>
Effective only:
●
For the R&S
NRP LAN power sensors.
●
SYSTem:COMMunicate:NETWork:IPADdress:MODE
is set to
STATic
.
Sets the IP address of the default gateway.
Configuring the system