Operating basics
R&S
®
NGU
42
User Manual 1179.2531.02 ─ 03
V
out
V
set
Voltage regulation
I
out
I
max
Cu
rr
en
t c
on
tr
ol
Figure 5-11: Current limit
CC mode
The current I
max
corresponds to the current setting adjustable in the instrument.
If I
out
reaches I
max
, the instrument switches to CC mode, i.e. the output current remains
constant and limited to I
max
even if the load increases. Instead, the output voltage V
out
decreases below V
set
. In a short circuit, the output voltage drops towards zero. In CC
mode, the font text in the channel display area changes to red.
See
and
Chapter 6.8.1, "Overcurrent protection (OCP)"
CR mode
To go into CR mode, the R&S
NGU must operate in sink mode where current flows into
the instrument and "Constant Resistance" are enabled.
With "Constant Resistance" enabled and configured, the R&S
NGU can set the resist-
ance in sink mode, this allows the R&S
NGU to behave like an electronic load which is
useful in battery test application.
In CR mode, the font text in the channel display area changes to cyan.
See
.
5.5.1
Voltage and current priority modes
By default, the R&S
NGU uses voltage priority mode, VPM to regulate output. In this
mode, the output is controlled by a bipolar voltage control loop. In VPM, the output cur-
rent can be set to the required positive and negative limits.
Operation modes