Remote Control Commands
R&S
®
NGL200/NGM200
100
User Manual 1178.8736.02 ─ 06
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit:LOWer?
Sets or queries the lower safety limit for voltage.
Setting parameters:
<voltage>
numeric | MIN | MAX
numeric
Numeric value for safety limit.
MIN
Min value for lower safety limit.
MAX
Max value for lower safety limit.
Range:
0.000E+00 to 2.050E+01
Increment: 0.001
*RST:
0.000E+00
Example:
See
Example "Configuring the output voltage"
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]
<voltage>
[SOURce:]VOLTage[:LEVel][:IMMediate]:ALIMit[:UPPer]?
Sets or queries the upper safety limit for voltage.
Setting parameters:
<voltage>
numeric | MIN | MAX
numeric
Numeric value for upper safety limit.
MIN
Min value for upper safety limit.
MAX
Max value for upper safety limit.
Range:
0.000E+00 to 2.050E+01
Increment: 0.001
*RST:
2.050E+01
Example:
See
Example "Configuring the output voltage"
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer
<current>
[SOURce:]CURRent[:LEVel][:IMMediate]:ALIMit:LOWer?
Sets or queries the lower safety limit for current.
Setting parameters:
<current>
numeric | MIN | MAX
numeric
Numeric value for lower safety limit.
MIN
Min value for lower safety limit.