Instrument Functions
R&S
®
NGL200
42
User Manual 1178.8736.02 ─ 02.01
Figure 6-4: Constant resistance dialog
6.2.2
Fast Transient Response
With fast transient response, the R&S
NGL200 is able to quickly stabilize the output
voltage upon a step change in the load current.
Load transient recovery time is between 30
μ
s or 100
μ
s for the output voltage to
recover and stay within 20 mV of the nominal output voltage follows by a step change
in the load current, typically equal to the full load current rating of the supply.
Figure 6-5: Transient response graph
1. Press [Settings] key.
The R&S
NGL200 displays the main menu window.
2. Select the required channel tab to apply fast transient response.
The R&S
NGL200 displays the selected channel menu.
3. Activate the "Fast Transient Response" from the menu.
Activating the Channels Output