Instrument Functions
R&S
®
NGL200
41
User Manual 1178.8736.02 ─ 02.01
Figure 6-3: Output of Ch2 in CV mode
Multiple outputs can be turned on or off at the same time.
See also
6.2.1
Set Constant Resitance
By enabling the constant resistance (CR) mode, you can operate the R&S
NGL200 as
an electronic load in sink mode. This allows you to perform testing that requires a con-
stant load resistor in your application.
1. Press [Settings] key.
The R&S
NGL200 displays the main menu window.
2. Select the required channel tab to apply constant resistance.
The R&S
NGL200 displays the selected channel menu.
3. Select "Constant Resistance" from the menu.
The R&S
NGL200 displays the "Constant Resistance" dialog.
4. Activate the "Enabled" menu item.
When operates in sink mode condition, the R&S
NGL200 displays the "Constant
Resistance" icon on the selected channel status bar information.
5. Set the required resistance.
The R&S
NGL200 displays an on-screen keypad to set the value.
6. Confirm value with the unit key (m
Ω
or
Ω
).
Activating the Channels Output