Remote Control Commands
R&S
®
NGL200
125
User Manual 1178.8736.02 ─ 02.01
LOG:DURation
<span>
LOG:DURation?
[<span>]
Sets/Queries the duration of the data logging.
Setting parameters:
<span>
numeric_values | MIN | MAX
numeric_values
Duration of the data logging captured in the range of 0 s to
3.49*10^5 s.
MIN
Minimum duration of the data logging captured at 0 s.
MAX
Maximum duration of the data logging captured at 3.49*10^5 s.
Default unit: s
Query parameters:
<span>
MIN | MAX
Return the duration of the data logging.
Example:
LOG:DUR MAX
LOG:DUR? MAX -> 349000
LOG:FNAMe
<filename>[,<loaction>]
LOG:FNAMe?
[<location>]
Sets/Queries the filename and storage location for the logging function.
Setting parameters:
<filename>
Filename of the data logging function.
<location>
INT | EXT | DEF
INT
Internal memory.
EXT
USB stick.
DEF
Internal memory.
Query parameters:
<location>
INT | EXT | DEF
Storage location for the logging function.
Example:
LOG:FNAM “Test01.CSV“,INT
LOG:FNAM? -> "/INT/DATA/Test01.CSV"
Data and File Management Commands