Remote Control
R&S
®
NGE100B
85
User Manual 5601.1343.02 ─ 04
3
channel Ch 3
Example:
INST OUT1
FUSE:LINK 2
FUSE:UNL 2
Fuse Ch 1 is unlinked with Ch 2
Usage:
Setting only
FUSE[:STATe]
<arg0>
This command defines or queries the fuse for the previous selected channel.
Parameters:
<arg0>
{0 | 1}
1
Fuse will be activated.
0
Fuse will be deactivated.
*RST:
OFF | 0
Example:
INST OUT1
FUSE ON
Fuse of Ch 1 is activated
FUSE?
-> 1
Return fuse state of the selected channel.
6.5.8
OVP Setting
[SOURce:]VOLTage:PROTection:MODE
<arg0>
This command defines or queries the OVP mode for the previous selected channel.
Parameters:
<arg0>
MEASured | PROTected
MEASured
The OVP switches off if the measured value exceeds the thresh-
old.
PROTected
If the adjusted threshold is exceeded the output of the instru-
ment will be not switched on; additionally the measured value is
monitored (please also refer to function MEASured).
Example:
INST OUT1
VOLT:PROT:MODE PROT
Sets OVP protected mode for Ch 1.
VOLT:PROT:MODE?
-> PROT
Return the selected channel OVP mode.
Remote Control - Commands