Remote control commands
R&S
®
LCX Series
177
User Manual 1179.2260.02 ─ 02
SHORt
Sets the measurement time
≤
.015
s.
MEDium
Sets the measurement time
≤
0.100
s.
LONG
Sets the measurement time
≤
0.500
s.
DEFault
Uses the default setting "SHORt".
Example:
See
Example "Setting key parameters"
Manual operation:
See
CURRent[:LEVel]
<Current Level>
CURRent[:LEVel]?
<Current Level>
Sets the test signal current in RMS (root mean square).
Parameters for setting and query:
<Current Level>
Range:
0.0001 A to 0.2 A
Default unit: A
Example:
See
Example "Setting key parameters"
Manual operation:
See
FREQuency[:CW]
<Test Signal Frequency>
Sets the frequency of the test signal.
Parameters:
<Frequency>
float
Range:
4 to 300 * 10
3
(depends on instrument model and
options)
*RST:
1E+3 Hz
Default unit: Hz
Example:
See
Example "Setting key parameters"
Manual operation:
See
VOLTage[:LEVel]
<Voltage Level>
Sets the test signal voltage in RMS (root mean square).
Parameters:
<Voltage Level>
<numeric>
Sets the value.
MIN | MINimum
Queries the lower limit of the signal level.
Test signal commands