
Remote Commands
R&S
®
FSW
984
User Manual 1173.9411.02 ─ 43
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:PHOLd:RESult?
This command queries the maximum RMS of the time domain power. The query is only
possible if the peak hold function has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:PHOLd
.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<RMSPower>
RMS power of the signal during the measurement time.
Usage:
Query only
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:RMS:RESult?
This command queries the RMS of the time domain power.
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Suffix:
<n>
.
<m>
Return values:
<RMSPower>
RMS power of the signal during the measurement time.
Usage:
Query only
Manual operation:
See
Measuring the Standard Deviation
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:SDEViation:AVERage:RESult?
This command queries the average standard deviation of the time domain power. The
query is only possible if averaging has been activated previously using
CALCulate<n>:MARKer<m>:FUNCtion:SUMMary:AVERage
To get a valid result, you have to perform a complete measurement with synchroniza-
tion to the end of the measurement before reading out the result. This is only possible
for single sweep mode.
See also
Configuring and Performing Measurements