Remote Commands
R&S
®
FSW
930
User Manual 1173.9411.02 ─ 43
MAX
The maximum of the relative
and
the absolute level is used as
the limit start value. Use the
RANGe<ri>:LIMit<li>:RELative:STARt
]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:
STARt:ABS
commands to define these values.
OFF
No function is used, the relative limit line is defined by a fixed
relative start value. Use the
RANGe<ri>:LIMit<li>:RELative:STARt
define this value.
*RST:
OFF
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STARt:
FUNCtion MAX
Chapter 14.5.7.11, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP
<Level>
This command defines a relative limit for a SEM range.
Unlike manual operation, you can define a relative limit anytime and regardless of the
limit check mode.
Suffix:
<sb>
.
1 to 8
Sub block in a Multi-SEM measurement
<ri>
1..n
Selects the SEM range.
<li>
1..n
Power class for which the limit is defined.
Parameters:
<Level>
Relative limit at the stop frequency of a SEM range.
Range:
-400 to 400
*RST:
-50
Default unit: dBc
Example:
SENSe:ESPectrum:RANGe:LIMit:RELative:STOP -15
Chapter 14.5.7.11, "Example: SEM
Manual operation:
See
[SENSe:]ESPectrum<sb>:RANGe<ri>:LIMit<li>:RELative:STOP:ABS
<Level>
This command defines an absolute limit for the MAX function of the relative limit for a
SEM range.
Configuring and Performing Measurements